A methodology for assessing and improving the total geometric accuracy in gamma knife radiosurgery

 
see the original item page
in the repository's web site and access all digital files if the item*
share




2013 (EN)

A methodology for assessing and improving the total geometric accuracy in gamma knife radiosurgery

Georgiou, E.
Torrens, M.
Karavasilis, E.
Karaiskos, P.
Moutsatsos, A.
Pantelis, E.
Seimenis, I.
Pappas, E.

The purpose of the study is to present an end-to-end experimental procedure, based on a polymer gel phantom, capable of assessing the total geometric uncertainty in GK radiosurgery applications, in which MR images are solely used for both target delineation and registration of patient image coordinates to the Leksell space. As a result the study aims to propose a time-efficient method, based on corresponding polymer gel results, which considerably improves the geometric accuracy in GK treatment delivery.

Text
Text (Conference or workshop item (Presentation))

θεραπευτική
ιατρική
φυσική
πυρηνική ιατρική


English

2013-06-20


11th International Stereotactic Radiosurgery Society Congress, Toronto, Canada, 16-20 June 2013